IS specification Mark IS: 3829 (PART-3), WHO GMP
Doors :Double Hinged Door
Sterilizer Trolley:SS 304 two numbers with locking arrangement
Carriage : SS 304 one number
Sterilizer Material of Construction:
Chamber : SS 316
Jacket & Non-Contact Parts: SS 316
Insulation: 75-100 mm thick Mineral Wool
Piping: SS 304
Door: Made high Grade SS 304.
Door Sealant: High Grade Silicon Gasket
Finish: Internal: Mirror External: Matt
The equipment is smooth finished without any sharp edges and crevices.
Required Process Control and Other Features:
Machine Operation Control:By Programmable Logic Controller (PLC) make Allen Bradley
Machine operational interface:By Pre-programmed Coloured HMI 4.3 touch screen make Allen Bradley
1)The machine must support TCP/ IP protocol
2)Communication protocol: Through MODBUS TCP/ MODBUS ASCII/ RTU
Process specification:
Sterilizing Temperature: Minimum: 121C -Maximum: 134C
Sterile Hold Time: Minimum: 15 min.
Maximum: As required
Chamber Pressure at Sterilization: Minimum: 1.12 bar-Maximum: 2.1 bar
The Machine is capable of running in automatic mode via touch screen human machine interface (HMI)
Machine air filter is capability to remove 99.97 % pathogens.
Machine has 3 pre-programmed cycles.
The machine has user variable cycle where user can customize the temperature settings, pressure settings and vacuum pulse settings.
The sterilizer setting is password protected.
The date and time setting is password protected and User is able to update date and time settings on the screen for better record keeping.
The machine is equipped with water detection system. The user will be notified in case the water level is not optimum for regular machine operation.
The Sterilizer has automatic water feeding system.
The user is record the progress of the sterilization cycle in real time therefore online real time printing facility will be provided.
The user cn print the following details:
Date and time of the cycle.Status of the cycle.
Temperature at drain and chamber pressure at every event change and during sterile hold it can be printed at desired interval, minimum of 1 minute.
All undesired situation (faults) are to be printed at the time of fault generation.
The Sterilizer has Hinged Door with shooting bolts locking system.
Door frame is strong & robust enough to ensure smooth door movement & consistency in alignment.
Door movement is smooth to ensure proper door alignment over a long period of time.
Door opening safety:
Door will be open only at the end of cycle and only when chamber is at room pressure and desired chamber temperature is reached.
Insulation: All the external surfaces are insulated (except where insulation would interfere with the function and operation of the sterilizer), to minimize the heat transmission to the environment.
The machine is equipped with air filter which removes 99.95% of the particulate matter of size 0.3 m so the drying does not compromise the sterility of the sterilized load.
![]() |